Issue StoriesNitric Oxide Delivery Via Nasal Cannulaby Kathleen M. Boyle, MS, RRT Delivery of nitric oxide via nasal cannula in nonintubated patients is safe and efficient, which helps fight the negative effects of cardiopulmonary hypertension In the scheme of things medical, nitric oxide is a newcomer to the armamentarium. It was identified in 1987 as the substance that produces relaxation of the endothelial tissue in the body (and known as endothelial-derived relaxing factor from 1980 until 1987).1 The first fledgling experiments were done in 1989, and research activity continues today. Nitric oxide was declared the molecule of the year2 by Science in 1992, and one vendors version of inhaled nitric oxide was approved by the US Food and Drug Administration (FDA) in 1999 for use in infants (of more than 34 weeks gestational age) with persistent pulmonary hypertension of the newborn (PPHN).3 The FDA-approved inhaled nitric oxide is a gaseous blend of 0.8% nitric oxide and 99.2% nitrogen (which is used for stabilization of the gas).3 Inhaled nitric oxide remains an experimental therapy for all patients who do not fall into the approved category (newborns with PPHN). Many cells in the body produce nitric oxide. Nitric oxide activates guanylate cyclase, which synthesizes a second-messenger molecule, cyclic guanosine monophosphate (cGMP). In smooth-muscle cells, the increased cGMP results in a reduction in intracellular calcium and a subsequent relaxation of the smooth muscle. Once nitric oxide enters the intravascular space, it binds to hemoglobin, forming nitrosyl iron (II) hemoglobin and then methemoglobin. Eventually, nitric oxide is converted to nitrates and nitrites in plasma and excreted, primarily by the kidney. The half-life of nitric oxide is 41 seconds,4 and since it binds to hemoglobin and becomes inactive, it has a local effect. Thus, inhaled nitric oxide is known as a potent pulmonary vasodilator. Delivery via Nasal Cannula
To ensure that this was a safe way to deliver nitric oxide, ambient nitric oxide and nitrogen dioxide levels were studied.5 These ambient nitric oxide and nitrogen dioxide levels were created through the simulated administration of nitric oxide via nasal cannula. This involved running 40 ppm of nitric oxide through the delivery system in an empty patient room for 8 hours. Two environmental-airsampling tubes were placed in the room, 61 cm and 1.52 m from the nasal prongs. The sampling tubes were replaced at hour four. The oxygen flow rate was 2.5 L/min and the nitric oxide flow rate was less than 100 mL/min. The levels of nitric oxide and nitrogen dioxide in the tubes were less than the detection limit of the laboratory (0.3 ppm). Therefore, nitric oxide levels did not exceed the US Occupational Safety and Health Administration (OSHA) permissible exposure limit of 25 ppm or the American Conference of Governmental Industrial Hygienists (ACGIH) threshold limit value of 25 ppm.5 In addition, nitrogen dioxide levels did not exceed the OSHA permissible exposure limit of 5 ppm or the ACGIH threshold limit value of 3 ppm.5 Nitric Oxide in Nonintubated Patients Questions often arise regarding oxygen delivery for these patients. At ACH, we use nitric oxide via nasal cannula with the addition of oxygen via nonrebreathing mask if the patient requires more oxygen than is delivered via the nasal cannula. Using a nonrebreathing mask will prevent any exhaled nitric oxide or nitrogen dioxide from entering the reservoir bag. If a partial-rebreathing mask were to be used, exhaled nitric oxide could enter the reservoir bag; this would allow time for the reaction with oxygen that creates nitrogen dioxide. This, plus the exhaled nitrogen dioxide already formed, could place a patient at risk for receiving a toxic dose of nitrogen dioxide. Nitric Oxide Via Nasal Cannula
A delivery system for inhaled nitric oxide, cleared by the FDA in early 2000 for US distribution, draws gas through a monitoring system at a rate of 230 mL/min. Nasal-cannula flow rates set at less than 500 mL/min will not consistently deliver gas to the patient. The system, therefore, had to be adapted for use with infant/pediatric and adult cannulae. Oxygen tubing is used to connect the nipple adapter of the flowmeter to the injector module. A 4.5 mm endotracheal-tube adapter and a 22 mm male/15 mm female x 22 mm male/15 mm female adapter connect the oxygen tubing to the injector module. A step-down adapter is used to connect the outlet side of the injector module to the nasal cannula. The nasal cannula is cut approximately 15 to 25 cm ahead of the bifurcation of the cannula leading to the nasal prongs. The cut ends of the cannula are connected to endotracheal-tube adapters (2.5 mm endotracheal-tube adapters for infant/pediatric cannulae and 3.5 mm endotracheal-tube adapters for adult cannulae). One of the adapters will fit into the nitric oxide sampling adapter, while a 22 mm male/15 mm female x 22 mm male/15 mm female adapter will connect the sampling adapter to the other section of the cannula and its endotracheal-tube adapter. This system has been used for the past 2 years without problems, but the systems manufacturer has released an application update stating that flow rates should not be less than 1 L/min.8 The update also indicates that increasing the inhaled nitric oxide setting may cause a bolus of nitrogen dioxide to be delivered to the patient. There is no evidence of this effect occurring with the system used at ACH. Cautions Nitrogen dioxide is a toxic gas. Among its effects are pulmonary edema, hemorrhage, changes in surfactant, and changes in the epithelium of the terminal bronchioles.9 OSHA has set safety limits for nitrogen dioxide at 5 ppm. Nitrogen dioxide is produced spontaneously from nitric oxide and oxygen.9 There is a potential for conversion of nitric oxide and oxygen into nitrogen dioxide within lung units if the residence time is more than 7 minutes.9 Inhaled nitric oxide itself can be toxic to the lungs at high concentrations. Nitric oxide levels of less than 100 ppm are believed to be safe for most patients, with close monitoring of nitrogen dioxide and methemoglobin levels. Patients should be treated using the lowest level of nitric oxide that produces a response. Conclusion Kathleen M. Boyle, MS, RRT, is a research therapist, Respiratory Care Services Department, Arkansas Childrens Hospital, Little Rock. References |
|
|
Featured Jobs
Find a Job |
ADDITIONAL ONLINE RESOURCES |
Featured Employer
|